  Level Meter  Distortion Meter
ÚÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿
³--ÚÄÄÄÄÄÄÄÄÄÄÄ¿ÚÄÄÄÄÄÄÄÄÄÄÄ¿--³   I have been working on one of these
³  ³ú          ³³ú          ³  ³   old AF automatic distortion meters.
³  ³ '.        ³³ '.        ³  ³   It is quite different to the usual
³--ÀÄÄÄÄÄÄÄÄÄÄÄÙÀÄÄÄÄÄÄÄÄÄÄÄÙ--³   fiddly variable Wien Bridge deep
³                              ³   notch type. So it is ideally suited
³        0                     ³   for setting up tape machines, where
³     Function     Distort     ³   the playback level & frequency will
³     [][][][]    [][][][]     ³   never be that steady.
³ O                          o ³
³       [][][][][][][][]       ³
³ O                          í ³
ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÙ
 In        Level Range     Mains

Level Ranges:- 30V, 10V, 3V, 1V, 300mV, 100mV, 30mV, 10mV.
            :-  30,  20, 10,  0,   -10,   -20,  -30,  -40 dBV

Response    :-    10Hz - 250kHz @ -3dB

Distortion Filters 400Hz, 1kHz & External.

Distortion Ranges: 30%, 10%, 3%, 1% min to 0.2%
            & dBs -10 -20 -40 -40  min to -55dB

Harmonic Response: 1.5x Filter to 130kHz @ -3dB

Mains 230V. odd 3 pin 1cm square socket.

The automatic refers to the distortion measurement does not need the zero level
reference level to be set first. (But is must be reading 30-100% on the level
meter scale.) This is done not by a very low distortion AGC system as you might
think, but a cleaver meter movement with 2 coils & no return springs. One coil
is fed with rectified DC from the input level detector 1 & acts as the return
spring & the other is fed from an 2nd detector from a high pass filter that
removes all of the fundamental, so it just the harmonics. The resultant meter
deflection is the distortion fraction.

In ÚÄÄÄÄÄ¿ ÚÄÄÄÄÄ¿    ACÚÄÄÄÄÄ¿              ÚÄÄÄÄÄÄÄÄÄÄÄ¿
oÄÄ´LevelÃÄ´InputÃÄÂÄÄÄÄ´LevelÃÄÄÄÄÄÄÄÄÄÄÄÄÄÄ´LEVEL METERÃÄÄÄÄÄ¿
   ³Atten³ ³ Amp ³ ³    ³Det 1³ DC           ÀÄÄÄÄÄÄÄÄÄÄÄÙ     ³
   ÀÄÄÄÄÄÙ ÀÄÄÄÄÄÙ ³    ³     ÃÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ¿ ³
                   ³    ÀÄÄÄÄÄÙ                         ÚÄÄÄÄÁÄÁÄÄÄÄÄ¿
                   ³   ÚÄÄÄÄÄÄÄÄÄ¿                      ³ DISTORTION ³
                   ³ oÄ´400Hz HPFÃÄo  ÚÄÄÄÄÄ¿  ÚÄÄÄÄÄ¿  ³2 COIL METER³
                   ³   ÃÄÄÄÄÄÄÄÄÄ´    ³ Set ³  ³Level³  ÀÄÄÄÄÂÄÂÄÄÄÄÄÙ
                   À>oÄ´ 1kHz HPFÃÄo<Ä´Gain ÃÄÄ´Det 2ÃÄÄÄÄÄÄÄÙ ³
                       ÀÄÄÄÄÄÄÄÄÄÙ    ³ Amp ³  ³     ÃÄÄÄÄÄÄÄÄÄÙ
                     oÄ External ÄÄo  ÀÂÂÂÂÂÙ  ÀÄÄÄÄÄÙ DC
                                        ^ÄÄÙDist Range

It just uses simple RC transistor Amps & bridge diodes for the electronics. The
notches are sealed LC high Q units. (M derived?)

There are only 3 presets:- Level gain, Distortion gain, & input amp Bias.

DISTORTION MEASURING PRINCIPLES
                                      f1  f2
     ³  f1                     0dBÄ´   |  ÚÁÄÄÄÄÄÄ     ³  f1
Level³   ³  f2                     ³   |  ³ High       ³      f2
     ³   ³   ³  f3                 ³   |  ³ Pass       ³       ³  f3
     ³   ³   ³   ³  f4  f5   -60dB_³___|_/  Filter     ³       ³   ³  f4  f5
     ÀÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁ         ÀÄÄÄÄÄÄÄÄÄÄÄÄÄÄ     ÀÄÄÄÄÄÄÄÁÄÄÄÁÄÄÄÁÄÄÄÁ
       Distorted Signal                 Filter            Filtered Signal

The high pass filter must completely remove the fundamental & all LF hums etc.
but leave all the harmonics intact. The detectors (especially the distortion
one) should be a true RMS type, as the waveform will generally be very complex
& nothing like a sine wave. The ratio of f1 to all the harmonics f2+++ is the
distortion factor.

Calibration is done by adding in a 2nd tone (e.g. 3-10x freq) via high value Rs
at say 10% of the level to indicate 10% distortion. The distortion meter
bandwidth can also be checked with this circuit.

1V 1kHz Äo-oÄÄÄ10KÄÄÂÄÄÄÂÄÄÄÄÄÄ¿
 Osc 1   o          ³   ³      ³
        _³_         ³   ³   Distortion
                    ³  DVM    Meter
100mV               ³  _³_    _³_
2-20kHz Äo-oÄÄÄ10kÄÄÙ
 Osc 2   o
        _³_

To check the HPF filter is working OK a very pure sine wave is needed, so an LC
tuned circuit after a signal generator is used. But I did find that > 0.5V @
resonance on the meter did make the inductor distortion rise > 0.01% !

  Osc         88mH      500mV
50mV 1kHz ÄÄÄ((((()ÄÄÂÄÄÄÄÄ¿ Hi Z
 low Z               ³  Distortion
    ³        0.28uF ===   Meter
   _³________________³_____³

MODERN ANSWER
Of course with a good quality PC sound card & a Furrier analysis program that
can add up the harmonics distortion measurement should be easy.  


See my bul "Marconi 2331A Distortion Factor" & "Simple 1kHz AF osc".